Using the Calibration Programs
9-17
9.3.4 Manual Calibration Program
Running the calibrator
At the “
Count
” screen, run a calibration material with known expected results 11 consecutive
times as instructed by
Chapter 6 Operating Your Analyzer.
Calculating the new calibration factors manually
Press [MENU] to enter the system menu.
Figure 9-20 System menu
SELECT
“
Calibration
→
Manual
” (Figure 9-20) to enter the “
Manual
” screen ( Figure 9-21).
Figure 9-21 “Manual” screen
The left of the “
Manual
" screen displays the available calibration modes – “
Whole blood
”
and “
Prediluted
”. The right of the “
Manual
” screen displays the calibration factors of WBC,
RBC, HGB, MCV, PLT and the time the factors are saved.
Use the following formula to calculate the new calibration factor.
mean
recorded
result
ected
factor
old
factor
new
exp
×
=
Example:
Assuming for a certain calibrator, the expected WBC value (namely the reference value
Содержание BC-2800
Страница 1: ...BC 2800 Auto Hematology Analyzer Operator s Manual ...
Страница 2: ......
Страница 12: ......
Страница 24: ...Using This Manual 1 12 Figure 1 4 High voltage warning label 7 High Voltage 7 ...
Страница 62: ......
Страница 91: ...Customizing the Analyzer Software 5 29 Figure 5 46 Saving changes ...
Страница 92: ......
Страница 122: ...Operating Your Analyzer 6 30 z Dispose of reagents waste samples consumables etc according to government regulations ...
Страница 170: ...Using the QC Programs 8 14 Figure 8 18 Transmission dialog box ...
Страница 196: ......
Страница 248: ......
Страница 252: ......
Страница 258: ...Appendices B 6 38 6 cm 32 2 cm 43 7 cm B 13 Weight Less than 23 kg B 14 Contraindications None ...
Страница 266: ......
Страница 284: ...Appendices E 10 Hemoglobin Concentration ...
Страница 286: ...P N 2800 20 28795 2 0 ...