Troubleshooting Your Analyzer
11-5
6. If the difference between the reference
WBC count time and the actual WBC
count time is less than 2 seconds, the
error has been removed;
7. If the difference is still greater than 2
seconds but consistent, enter the
“
Setup
→
Settings
→
Count
” and
reset the WBC count time. Then enter
the “
Service
→
Self-test
” screen and
test the actual WBC count time as
instructed by
Chapter 10.5.1
to
confirm the difference is less than 2
seconds;
8. If the problem remains, contact
Mindray customer service department
or your local distributor.
WBC bubbles
1. Diluent or rinse running
out;
2. Loose
tube
connections;
3. Inappropriate
WBC
count time setting.
1. Check if the diluent or rinse has run
out. If so, change a new container of
diluent or rinse as instructed in
Chapter 4.4.1
;
2. Check the connection of the diluent
and rinse pickup tube. If necessary,
reconnect and tighten them as
instructed by
Chapter 4.4.1
;
3. If the problem remains, adjust the
WBC count time as instructed by
Chapter 5.3.5
;
4. If the problem remains, contact
Mindray customer service department
or your local distributor.
RBC clog
1. Clogged
the
aperture;
2. Inappropriate RBC
count time setting;
3. Solenoid valve error.
1. Enter the “
Service
→
Maintenance
”
screen. Zap and flush the aperture as
instructed by
Chapter 10.3.4.
and
10.3.5.
;
2. Enter the “
Setup
→
Settings
→
Count
” screen and record the RBC
count time. Then enter the “
Service
→
Self-test
” screen and test the actual
RBC count time as instructed by
Chapter 10.5.1
.;
3. If the difference between the reference
Содержание BC-2800
Страница 1: ...BC 2800 Auto Hematology Analyzer Operator s Manual ...
Страница 2: ......
Страница 12: ......
Страница 24: ...Using This Manual 1 12 Figure 1 4 High voltage warning label 7 High Voltage 7 ...
Страница 62: ......
Страница 91: ...Customizing the Analyzer Software 5 29 Figure 5 46 Saving changes ...
Страница 92: ......
Страница 122: ...Operating Your Analyzer 6 30 z Dispose of reagents waste samples consumables etc according to government regulations ...
Страница 170: ...Using the QC Programs 8 14 Figure 8 18 Transmission dialog box ...
Страница 196: ......
Страница 248: ......
Страница 252: ......
Страница 258: ...Appendices B 6 38 6 cm 32 2 cm 43 7 cm B 13 Weight Less than 23 kg B 14 Contraindications None ...
Страница 266: ......
Страница 284: ...Appendices E 10 Hemoglobin Concentration ...
Страница 286: ...P N 2800 20 28795 2 0 ...