Operating Your Analyzer
6-17
Figure 6-11 “Re-count” screen
Follow the previously introduced procedure to re-analyze the sample in question. The new
result will overwrite the old result while the sample information keeps unchanged.
Automatic saving of analysis results
This analyzer automatically saves a maximum of 10,000 sample results. When the maximum
number has been reached, the newest result will overwrite the oldest.
Parameter flags
If the analysis result is followed by an ”
H
” or “
L
”
,
it means the analysis result has
exceeded the upper or lower limit of the reference range.
If you see
***
as opposed to the result, it means the result is either unreliable or out of the
operating range.
If the WBC result is less than 0.5
×
10
9
/L, this analyzer will not perform the differential
analysis and all the related parameter values will be non-numeric (
***
)
.
z
The result of the background check will not be flagged.
Histogram flags
The system will flag abnormal histograms.
Abnormal WBC histograms will be flagged by one of the markings: R1, R2, R3, R4 and
R
m.
R
1
: indicates abnormality on the left side of the lymphocyte hump and possible presence of
Содержание BC-2800
Страница 1: ...BC 2800 Auto Hematology Analyzer Operator s Manual ...
Страница 2: ......
Страница 12: ......
Страница 24: ...Using This Manual 1 12 Figure 1 4 High voltage warning label 7 High Voltage 7 ...
Страница 62: ......
Страница 91: ...Customizing the Analyzer Software 5 29 Figure 5 46 Saving changes ...
Страница 92: ......
Страница 122: ...Operating Your Analyzer 6 30 z Dispose of reagents waste samples consumables etc according to government regulations ...
Страница 170: ...Using the QC Programs 8 14 Figure 8 18 Transmission dialog box ...
Страница 196: ......
Страница 248: ......
Страница 252: ......
Страница 258: ...Appendices B 6 38 6 cm 32 2 cm 43 7 cm B 13 Weight Less than 23 kg B 14 Contraindications None ...
Страница 266: ......
Страница 284: ...Appendices E 10 Hemoglobin Concentration ...
Страница 286: ...P N 2800 20 28795 2 0 ...