Maintaining Your Analyzer
10-15
Figure 10-19 Bath
5.
SELECT
“
Lyse test
”. The analyzer will automatically drain the bath and then dispense
2ml of lyse into the bath;
6. Check the scale to see whether the lyse has reached the expected line (the second from
the bottom). If so, press [ENTER] and the analyzer will automatically flush the bath and
dispense lyse and the test is done;
7. If not, repeat steps 5 and 6 several times. If all the tries have failed, check whether the
lyse has run out or the lyse pickup tube is not properly connected to this analyzer. If the
lyse is still enough and the tube is well connected to the analyzer, contact the Mindray or
your local distributor for assistance.
10.3.9 Clean Bath
Follow the steps given below to perform the “
Clean bath
” procedure:
SELECT
“
Clean bath
” to start the cleaning procedure, as Figure 10-20 shows. When the
cleaning is done, the screen will return to the initial state;
Содержание BC-2800
Страница 1: ...BC 2800 Auto Hematology Analyzer Operator s Manual ...
Страница 2: ......
Страница 12: ......
Страница 24: ...Using This Manual 1 12 Figure 1 4 High voltage warning label 7 High Voltage 7 ...
Страница 62: ......
Страница 91: ...Customizing the Analyzer Software 5 29 Figure 5 46 Saving changes ...
Страница 92: ......
Страница 122: ...Operating Your Analyzer 6 30 z Dispose of reagents waste samples consumables etc according to government regulations ...
Страница 170: ...Using the QC Programs 8 14 Figure 8 18 Transmission dialog box ...
Страница 196: ......
Страница 248: ......
Страница 252: ......
Страница 258: ...Appendices B 6 38 6 cm 32 2 cm 43 7 cm B 13 Weight Less than 23 kg B 14 Contraindications None ...
Страница 266: ......
Страница 284: ...Appendices E 10 Hemoglobin Concentration ...
Страница 286: ...P N 2800 20 28795 2 0 ...