Maintaining Your Analyzer
10-35
Figure 10-52 Removing screws
8. Remove the probe from the probe wipe and insert the localizer into the probe wipe from
the bottom, as Figure 10-53 shows;
Figure 10-53 Using localizer
9. Insert the probe into the probe wipe until it reaches the localizer, as Figure 10-54 shows;
Figure 10-54 Inserting sample probe into probe wipe
10. Retighten the screw to fix the probe and remove the localizer to wrap up the adjustment.
Содержание BC-2800
Страница 1: ...BC 2800 Auto Hematology Analyzer Operator s Manual ...
Страница 2: ......
Страница 12: ......
Страница 24: ...Using This Manual 1 12 Figure 1 4 High voltage warning label 7 High Voltage 7 ...
Страница 62: ......
Страница 91: ...Customizing the Analyzer Software 5 29 Figure 5 46 Saving changes ...
Страница 92: ......
Страница 122: ...Operating Your Analyzer 6 30 z Dispose of reagents waste samples consumables etc according to government regulations ...
Страница 170: ...Using the QC Programs 8 14 Figure 8 18 Transmission dialog box ...
Страница 196: ......
Страница 248: ......
Страница 252: ......
Страница 258: ...Appendices B 6 38 6 cm 32 2 cm 43 7 cm B 13 Weight Less than 23 kg B 14 Contraindications None ...
Страница 266: ......
Страница 284: ...Appendices E 10 Hemoglobin Concentration ...
Страница 286: ...P N 2800 20 28795 2 0 ...