Using the Calibration Programs
9-13
If all the parameter values obtained are numeric, a message box will pop up to confirm the
validity of the results, as Figure 9-14 shows.
CLICK
“
Yes
” to save the results to the “
Fresh
Blood
” screen;
CLICK
“
No
” to abort the result.
Figure 9-14 A message box to confirm the validity
Repeat the above steps to run the sample 3 to 5 times (5 is recommended) and the analyzer
will automatically calculate the CV and calibration factor, as shows. Note that the CVs should
be within the ranges specified in Table 9-1.
Figure 9-15 Calibration with fresh blood
The calculated calibration factors should be within the 75
%
to 125
%
. Any calculated value
that falls out of the calibration range will be flagged with a “*” at the upper right corner. Other
values will not be displayed. In case of an empty calibration factor, try to find out the reason
and if necessary, contact Mindray customer service department or your local distributor.
Содержание BC-2800
Страница 1: ...BC 2800 Auto Hematology Analyzer Operator s Manual ...
Страница 2: ......
Страница 12: ......
Страница 24: ...Using This Manual 1 12 Figure 1 4 High voltage warning label 7 High Voltage 7 ...
Страница 62: ......
Страница 91: ...Customizing the Analyzer Software 5 29 Figure 5 46 Saving changes ...
Страница 92: ......
Страница 122: ...Operating Your Analyzer 6 30 z Dispose of reagents waste samples consumables etc according to government regulations ...
Страница 170: ...Using the QC Programs 8 14 Figure 8 18 Transmission dialog box ...
Страница 196: ......
Страница 248: ......
Страница 252: ......
Страница 258: ...Appendices B 6 38 6 cm 32 2 cm 43 7 cm B 13 Weight Less than 23 kg B 14 Contraindications None ...
Страница 266: ......
Страница 284: ...Appendices E 10 Hemoglobin Concentration ...
Страница 286: ...P N 2800 20 28795 2 0 ...