Maintaining Your Analyzer
10-6
You should perform the “
Rinse prime
” procedure to prime the rinse tubing when
there are bubbles in the tubing; or
the rinse in the tubing is contaminated; or
you have installed a new container of rinse without shutting the analyzer.
At the “
Maintenance
” screen,
SELECT
“
Rinse prime
” to prime the tubing and the priming
progress will be displayed at the bottom of the screen, as Figure 10-4 shows. When the
priming is done, the screen will return to the initial state.
Figure 10-4 Priming rinse screen
10.3.3 Lyse Prime
z
Samples, controls, calibrator and waste are potentially infectious. Wear
proper personal protective equipment (e.g. gloves, lab coat, etc.) and follow
safe laboratory procedures when handling them in the laboratory.
z
After installing new reagents, let the reagents stand for a while before using
them.
z
After installing a new container of diluent, rinse or lyse, do a background
check to ensure the background results are normal.
Содержание BC-2800
Страница 1: ...BC 2800 Auto Hematology Analyzer Operator s Manual ...
Страница 2: ......
Страница 12: ......
Страница 24: ...Using This Manual 1 12 Figure 1 4 High voltage warning label 7 High Voltage 7 ...
Страница 62: ......
Страница 91: ...Customizing the Analyzer Software 5 29 Figure 5 46 Saving changes ...
Страница 92: ......
Страница 122: ...Operating Your Analyzer 6 30 z Dispose of reagents waste samples consumables etc according to government regulations ...
Страница 170: ...Using the QC Programs 8 14 Figure 8 18 Transmission dialog box ...
Страница 196: ......
Страница 248: ......
Страница 252: ......
Страница 258: ...Appendices B 6 38 6 cm 32 2 cm 43 7 cm B 13 Weight Less than 23 kg B 14 Contraindications None ...
Страница 266: ......
Страница 284: ...Appendices E 10 Hemoglobin Concentration ...
Страница 286: ...P N 2800 20 28795 2 0 ...