Using the Calibration Programs
9-12
Figure 9-12 Editing reference value
ENTER
the reference value into the “
Mean
” edit box. To correct any erroneous entry,
MODIFY
the digit. After you have finished the editing, press [F2] to exit the editing state.
Running the sample
After you have finished editing the calibration settings of Sample 1, refer to the sample
handling and analysis procedures introduced in
Chapter 6 Operating Your Analyzer
and
prepare the fresh blood samples in the selected count mode to perform the fresh blood
calibration.
Saving calibration results
If non-numeric parameter values (“***”) are obtained, a message box will pop up to warn you,
as Figure 9-13 shows.
Figure 9-13 A message box to warn you about the invalid results