Troubleshooting Your Analyzer
11-4
5.3.4
to set the voltage within 3.4 to
4.8V, preferably 4.5V;
3. If the problem remains, shut down
your analyzer and contact Mindray
customer service department or you
local distributor.
HGB adjust
HGB blank voltage within 3.2
V to 3.4 V or 4.8 V to 4.9 V.
1. Do the probe cleanser cleaning
procedure as instructed in
Chapter
10.3.6.
;
2. If the problem remains, adjust the
HGB gain as instructed by
Chapter
5.3.4
to set the voltage within 3.4 to
4.8V, preferably 4.5V;
3. If the problem remains, shut down
your analyzer and contact Mindray
customer service department or you
local distributor.
WBC clog
1. Clogged
the
aperture;
2. Inappropriate WBC count
time setting;
3. Solenoid valve error.
1. Enter the “
Service
→
Maintenance
”
screen. Zap and flush the aperture as
instructed by
Chapter 10.3.4
and
10.3.5
;
2. Enter the “
Setup
→
Settings
→
Count
” screen and record the WBC
count time. Then enter the “
Service
→
Self-test
” screen and test the actual
WBC count time as instructed by
Chapter 10.5.1
.;
3. If the difference between the reference
WBC count time and the actual WBC
count time is less than 2 seconds, the
error has been removed;
4. If not, enter the “
Service
→
Maintenance
” screen and do the
probe cleanser cleaning procedure as
instructed by
Chapter 10.3.6
;
5. Enter the “
Setup
→
Settings
→
Count
” screen and record the WBC
count time. Then enter the “Service
→
Self Test” screen and test the actual
WBC count time as instructed by
Chapter 10.5.1
;
Содержание BC-2800
Страница 1: ...BC 2800 Auto Hematology Analyzer Operator s Manual ...
Страница 2: ......
Страница 12: ......
Страница 24: ...Using This Manual 1 12 Figure 1 4 High voltage warning label 7 High Voltage 7 ...
Страница 62: ......
Страница 91: ...Customizing the Analyzer Software 5 29 Figure 5 46 Saving changes ...
Страница 92: ......
Страница 122: ...Operating Your Analyzer 6 30 z Dispose of reagents waste samples consumables etc according to government regulations ...
Страница 170: ...Using the QC Programs 8 14 Figure 8 18 Transmission dialog box ...
Страница 196: ......
Страница 248: ......
Страница 252: ......
Страница 258: ...Appendices B 6 38 6 cm 32 2 cm 43 7 cm B 13 Weight Less than 23 kg B 14 Contraindications None ...
Страница 266: ......
Страница 284: ...Appendices E 10 Hemoglobin Concentration ...
Страница 286: ...P N 2800 20 28795 2 0 ...