Maintaining Your Analyzer
10-26
Figure 10-37 testing mechanic part
To conduct the following tests,
SELECT
the desired test and the results will be displayed
later.
Syringe
motor
The syringe motor controls the aspiration volume. This test checks whether the motor
functions normally.
Rotation
motor
The rotation motor rotates the sample probe inside the analyzer. This test checks whether the
motor functions normally.
Elevator
motor
The elevator motor controls elevation of the sample probe. This test checks whether the
motor functions normally.
This test checks whether the recorder or printer functions normally. If normal, when you press
[ENTER], the recorder or printer will print out a test page; if abnormal, the screen will display
the corresponding error message and you can see
Chapter 11 Troubleshooting
Your
Analyzer
for solutions.
10.5.3 Testing Valves
To test whether the valves function properly, press [F1] to select the “
Valve
” group, as Figure
10-38 shows. Follow the steps below to test the desired valve.
Содержание BC-2800
Страница 1: ...BC 2800 Auto Hematology Analyzer Operator s Manual ...
Страница 2: ......
Страница 12: ......
Страница 24: ...Using This Manual 1 12 Figure 1 4 High voltage warning label 7 High Voltage 7 ...
Страница 62: ......
Страница 91: ...Customizing the Analyzer Software 5 29 Figure 5 46 Saving changes ...
Страница 92: ......
Страница 122: ...Operating Your Analyzer 6 30 z Dispose of reagents waste samples consumables etc according to government regulations ...
Страница 170: ...Using the QC Programs 8 14 Figure 8 18 Transmission dialog box ...
Страница 196: ......
Страница 248: ......
Страница 252: ......
Страница 258: ...Appendices B 6 38 6 cm 32 2 cm 43 7 cm B 13 Weight Less than 23 kg B 14 Contraindications None ...
Страница 266: ......
Страница 284: ...Appendices E 10 Hemoglobin Concentration ...
Страница 286: ...P N 2800 20 28795 2 0 ...