Maintaining Your Analyzer
10-4
Totally 13 maintenance procedures are available at the “
Maintenance
” screen.
Diluent
Prime
Rinse
Prime
Lyse
Prime
Zap
Aperture
Flush
Aperture
Probe
Cleanser
Cleaning
E-Z Cleanser Cleaning
Lyse
Test
Clean
Bath
Drain
Bath
Drain
Tubing
Clean Probe Wipe
Prepare to Ship
10.3.1 Diluent Prime
z
The reagents are irritating to eyes, skin and diaphragm. Wear proper
personal protective equipment (e.g. gloves, lab coat, etc.) and follow safe
laboratory procedures when handling them in the laboratory.
z
If the reagents accidentally spill on your skin, wash them off with plenty of
water and if necessary, go see a doctor; if the reagents accidentally spill into
your eyes, wash them off with plenty of water and immediately go see a
doctor.
z
After installing new reagents, let the reagents stand for a while before using
them.
z
After installing a new container of diluent, rinse or lyse, do a background
check to ensure the background results are normal.
You should perform the “
Diluent prime
” procedure to prime the diluent tubing when
there are bubbles in the tubing; or
Содержание BC-2800
Страница 1: ...BC 2800 Auto Hematology Analyzer Operator s Manual ...
Страница 2: ......
Страница 12: ......
Страница 24: ...Using This Manual 1 12 Figure 1 4 High voltage warning label 7 High Voltage 7 ...
Страница 62: ......
Страница 91: ...Customizing the Analyzer Software 5 29 Figure 5 46 Saving changes ...
Страница 92: ......
Страница 122: ...Operating Your Analyzer 6 30 z Dispose of reagents waste samples consumables etc according to government regulations ...
Страница 170: ...Using the QC Programs 8 14 Figure 8 18 Transmission dialog box ...
Страница 196: ......
Страница 248: ......
Страница 252: ......
Страница 258: ...Appendices B 6 38 6 cm 32 2 cm 43 7 cm B 13 Weight Less than 23 kg B 14 Contraindications None ...
Страница 266: ......
Страница 284: ...Appendices E 10 Hemoglobin Concentration ...
Страница 286: ...P N 2800 20 28795 2 0 ...