Using the QC Programs
8-18
Press [PRINT] to print out all the settings.
Exiting the “
X-B Edit
” screen
Press [MENU] to exit the “
X-B Edit
” screen. a message box will pop up to remind you to save
the changes, as shows.
CLICK
“
Yes
” to save the changes;
CLICK
“
No
” to abort the changes.
Figure 8-24 A message box to confirm the changes
8.3.2 Running X-B Analysis
After you have enabled the X-B analysis and assigned valid means and ranges to the three
parameters, the system will automatically run the X-B analysis every 20 patient samples, of
which the MCV, MCH and MCHC results are within the set ranges.
8.3.3 Reviewing X-B Results
This analyzer automatically saves the results of the X-B analyses. You can review the saved
results in two modes - X-B table and X-B graph
X-B Table
Follow the steps introduced in
Chapter
8.3.1
.
X-B Graph
At the
“
X-B Table
” screen, press [F2] to enter the “
X-B Graph
” screen, as Figure 8-25 shows.
Содержание BC-2800
Страница 1: ...BC 2800 Auto Hematology Analyzer Operator s Manual ...
Страница 2: ......
Страница 12: ......
Страница 24: ...Using This Manual 1 12 Figure 1 4 High voltage warning label 7 High Voltage 7 ...
Страница 62: ......
Страница 91: ...Customizing the Analyzer Software 5 29 Figure 5 46 Saving changes ...
Страница 92: ......
Страница 122: ...Operating Your Analyzer 6 30 z Dispose of reagents waste samples consumables etc according to government regulations ...
Страница 170: ...Using the QC Programs 8 14 Figure 8 18 Transmission dialog box ...
Страница 196: ......
Страница 248: ......
Страница 252: ......
Страница 258: ...Appendices B 6 38 6 cm 32 2 cm 43 7 cm B 13 Weight Less than 23 kg B 14 Contraindications None ...
Страница 266: ......
Страница 284: ...Appendices E 10 Hemoglobin Concentration ...
Страница 286: ...P N 2800 20 28795 2 0 ...