Using the QC Programs
8-17
Enabling/disabling X-B analysis
Random samples are required for the X-B analysis. In case of known samples of a particular
type (oncology, neonatal and so forth) that will seriously interfere with the X-B results, disable
the X-B analysis.
SELECT
the combo box to the right of “
X-B analysis
”,
SELECT
“
On
”
or
“
Off
”
from
the
pull-down list
to enable or disable the X-B analysis, as Figure 8-23 shows.
Figure 8-23 Enabling/disabling X-B Edit
Entering the expected result (mean) and limit (range)
The expected results vary depending on laboratories. It is recommended they are obtained
by calculating the averages of at least 500 random patient samples. The recommended limit
is 5% to 10%.
z
Calibrate your analyzer before trying to establish the expected results by
calculating the averages of random patient samples.
ENTER
the expected mean and range respectively into the “
Mean
” box and “
Range
” boxes
of the parameters to be included in the QC run.
Deleting
settings
Press [DEL] to delete all the settings.
Printing
settings
Содержание BC-2800
Страница 1: ...BC 2800 Auto Hematology Analyzer Operator s Manual ...
Страница 2: ......
Страница 12: ......
Страница 24: ...Using This Manual 1 12 Figure 1 4 High voltage warning label 7 High Voltage 7 ...
Страница 62: ......
Страница 91: ...Customizing the Analyzer Software 5 29 Figure 5 46 Saving changes ...
Страница 92: ......
Страница 122: ...Operating Your Analyzer 6 30 z Dispose of reagents waste samples consumables etc according to government regulations ...
Страница 170: ...Using the QC Programs 8 14 Figure 8 18 Transmission dialog box ...
Страница 196: ......
Страница 248: ......
Страница 252: ......
Страница 258: ...Appendices B 6 38 6 cm 32 2 cm 43 7 cm B 13 Weight Less than 23 kg B 14 Contraindications None ...
Страница 266: ......
Страница 284: ...Appendices E 10 Hemoglobin Concentration ...
Страница 286: ...P N 2800 20 28795 2 0 ...