Using the Calibration Programs
9-10
Verifying new calibration factors
Press [MENU] to exit the “
Calibrator
” screen, a message box will pop up to confirm the new
calibration factors, as Figure 9-9 shows.
Figure 9-9 A message box to confirm the new calibration factors
CLICK
”
Yes
” to save the new calibration factors and enter the “
Count
” screen from the
system menu.
At the “
Count
” screen, run the calibrator or a normal-level control at least 3 consecutive times
and calculate the mean of the results. Compare the obtained means to the expected means.
Other operations
Printing new calibration factors
Press [PRINT] to print out the current calibration factors.
Exiting the ”
Calibrator
” screen
Press [MENU] to exit to system menu.
9.3.3 Fresh Blood Calibration
ENTER
the administrator password in the “
Password
” screen. Press [MENU] to enter the
system menu.
Figure 9-10 System menu
Содержание BC-2800
Страница 1: ...BC 2800 Auto Hematology Analyzer Operator s Manual ...
Страница 2: ......
Страница 12: ......
Страница 24: ...Using This Manual 1 12 Figure 1 4 High voltage warning label 7 High Voltage 7 ...
Страница 62: ......
Страница 91: ...Customizing the Analyzer Software 5 29 Figure 5 46 Saving changes ...
Страница 92: ......
Страница 122: ...Operating Your Analyzer 6 30 z Dispose of reagents waste samples consumables etc according to government regulations ...
Страница 170: ...Using the QC Programs 8 14 Figure 8 18 Transmission dialog box ...
Страница 196: ......
Страница 248: ......
Страница 252: ......
Страница 258: ...Appendices B 6 38 6 cm 32 2 cm 43 7 cm B 13 Weight Less than 23 kg B 14 Contraindications None ...
Страница 266: ......
Страница 284: ...Appendices E 10 Hemoglobin Concentration ...
Страница 286: ...P N 2800 20 28795 2 0 ...