Customizing the Analyzer Software
5-22
Setting reference ranges (administrators only)
1. Enter the administrator password as introduced in
Chapter 5.2.1
;
2. At the “
Settings
” screen, press [F1] to select the “
Ref. Range
” group, as Figure 5-35
shows;
3. When you have finished selecting the patient group,
SELECT
the desired parameter and
ENTER
the desired digits for the upper and lower limits;
Figure 5-35 Setting reference range
4. Press [F2] to save the changes. If the changes are successfully saved, a message box
shown in Figure 5-36 will pop up;
CLICK
the “
Yes
” to close the message box;
Figure 5-36 A message box to note data saved
5. If some entered limits are invalid, a message box shown in Figure 5-37 will pop up.
CLICK
“
Yes
” and enter valid number again;
Содержание BC-2800
Страница 1: ...BC 2800 Auto Hematology Analyzer Operator s Manual ...
Страница 2: ......
Страница 12: ......
Страница 24: ...Using This Manual 1 12 Figure 1 4 High voltage warning label 7 High Voltage 7 ...
Страница 62: ......
Страница 91: ...Customizing the Analyzer Software 5 29 Figure 5 46 Saving changes ...
Страница 92: ......
Страница 122: ...Operating Your Analyzer 6 30 z Dispose of reagents waste samples consumables etc according to government regulations ...
Страница 170: ...Using the QC Programs 8 14 Figure 8 18 Transmission dialog box ...
Страница 196: ......
Страница 248: ......
Страница 252: ......
Страница 258: ...Appendices B 6 38 6 cm 32 2 cm 43 7 cm B 13 Weight Less than 23 kg B 14 Contraindications None ...
Страница 266: ......
Страница 284: ...Appendices E 10 Hemoglobin Concentration ...
Страница 286: ...P N 2800 20 28795 2 0 ...