Maintaining Your Analyzer
10-36
10.10 Replacing the Probe Wipe
z
Samples, controls, calibrator and waste are potentially infectious. Wear
proper personal protective equipment (e.g. gloves, lab coat, etc.) and follow
safe laboratory procedures when handling them in the laboratory.
z
The sample probe tip is sharp and may contain biohazardous materials.
Exercise caution to avoid contact with the probe when working around it.
To replace the probe wipe:
1. Refer
to
Chapter 10.9
and do the steps 1 to 8;
2. Pull the loosen probe wipe upward to remove the probe wipe and disconnect the tubes
from the probe wipe (pay attention to the correspondence between the tubes and the
connectors), as Figure 10-55 shows;
Figure 10-55 Installing probe wipe
3. Install a new block and connect the tubing end with the black marking to the connector
below the block;
4. Calibration to the sample probe position should be performed after replacing the probe
wipe. Refer to
Chapter 10.9
.
Содержание BC-2800
Страница 1: ...BC 2800 Auto Hematology Analyzer Operator s Manual ...
Страница 2: ......
Страница 12: ......
Страница 24: ...Using This Manual 1 12 Figure 1 4 High voltage warning label 7 High Voltage 7 ...
Страница 62: ......
Страница 91: ...Customizing the Analyzer Software 5 29 Figure 5 46 Saving changes ...
Страница 92: ......
Страница 122: ...Operating Your Analyzer 6 30 z Dispose of reagents waste samples consumables etc according to government regulations ...
Страница 170: ...Using the QC Programs 8 14 Figure 8 18 Transmission dialog box ...
Страница 196: ......
Страница 248: ......
Страница 252: ......
Страница 258: ...Appendices B 6 38 6 cm 32 2 cm 43 7 cm B 13 Weight Less than 23 kg B 14 Contraindications None ...
Страница 266: ......
Страница 284: ...Appendices E 10 Hemoglobin Concentration ...
Страница 286: ...P N 2800 20 28795 2 0 ...