Appendix B
User’s Manual Probot
D960R3
B-7
MasterPressure.LowerLimit = 1.0
MasterPressure.UpperLimit = 400.0
Data_Collection_Rate =
2
TimeConstant =
0.1
Wait Sampler.Ready
InjectMode =
FullLoop
LowDispersionMode =
Off
UseAirSegment =
Off
UseHeadSpace =
Off
SyringeSpeed =
Normal
SyringeSpeedFactor =
0.2
SampleHeight =
5
FlushVolume =
5.0
WashVolume =
50
UV_VIS_1.Wavelength = 214
ColumnPressure.Formula Formula=Pump.Columnpressure.value
ColumnPressure.Type =
Analog
PumpPressure.Formula
Formula=Pump.Masterpressure.value
PumpPressure.Type =
Analog
0.000 UV.autozero
Flow =
2.50
%B = 0
%C =
0
%D =
0
Wait
UV.Ready and Pump.Ready and Sampler.Ready
Inject
UV_VIS_1.AcqOn
ColumnPressure.AcqOn
PumpPressure.AcqOn
Flow =
2.50
%B = 0
%C =
0
%D =
0
Probot_Start.On
Duration = 1.00
Trigger peakstart
UV_VIS_1.Delta > 0.1, true=2.00
Pump_Relay_1.State =
On
Pump_Relay_1.Duration = 1.00
EndTrigger
Trigger peakend
UV_VIS_1 < 1.0 and UV_VIS_1.Delta > -0.03, true=2.00
Pump_Relay_2.State =
On
Pump_Relay_2.Duration = 1.00
EndTrigger
17.500 %B =
50
18.000 %B =
50
18.100 %B =
0
25.000
UV_VIS_1.AcqOff
ColumnPressure.AcqOff
PumpPressure.AcqOff
Flow =
2.50
%B = 0
%C =
0
%D =
0
End
Содержание Probot 161403
Страница 1: ...D960R3 Probot TM Micro Fraction Collector User s Manual P N 161403 www lcpackings com ...
Страница 6: ...Table of Contents vi D960R3 User s Manual Probot This page intentionally left blank ...
Страница 10: ...Instructions for Returning Instruments x D960R3 User s Manual Probot This page intentionally left blank ...
Страница 14: ...CE Declaration xiv D960R3 User s Manual Probot This page intentionally left blank ...
Страница 68: ...Installation and Getting Started 2 48 D960R3 User s Manual Probot This page intentionally left blank ...
Страница 102: ...Appendix C C 6 D960R3 User s Manual Probot This page intentionally left blank ...
Страница 112: ...Appendix D D 10 D960R3 User s Manual Probot This page intentionally left blank ...