Installation and Getting Started
2-6
D960R3
User’s Manual Probot
2.4
Installing the Probot with the UltiMate Capillary HPLC System
The following section describes how to install the Probot Micro Fraction Collector
as a component in the LC Packings UltiMate Capillary HPLC system.
2.4.1
Electrical Connections
The Probot Micro Fraction Collector is controlled by the
µ
Carrier
™
software
package using a serial communication port of the personal computer (PC). In
addition, the Probot can be controlled by external events (e.g. START, TTL IN).
A manual control unit can be connected which allows for manual movement of
all three axes.
All electrical connections are made on the rear panel of the instrument
(FIGURE 1-4).
CAUTION
Caution: Avoid touching the electrical contacts on the terminal strips.
Electrostatic discharges could damage internal components. The manufacturer
will not accept any liability for damages directly or indirectly caused by
connecting the Probot Micro Fraction Collector to instruments which do not meet
relevant safety standards.
2.4.1 A
RS-232 Connector
Connect the RS-232 cable to the RS-232 connector on the Probot and to the
COM port of the personal computer.
TABLE 2-1
The Probot RS-232 Communication Cable
Probot (9-pin male)
PC (9-pin female)
Description
2 3 TxD
3 2 RxD
5 5 Ground
7 7 Reset
2.4.1 B
START Connector
If the external start feature is used, connect the Start Cable to the START
connector on the Probot and to the device that will provide the start signal for
the Probot (e.g. a TTL output of the UltiMate Micropump or a relay output of the
FAMOS).
TABLE 2-2
The START Cable
Input
Pin No.
Color Code
Description
Start 7
lilac
Start
input
Ground 9
black
Ground
connection
CAUTION
Caution: If the controlling device does not use a switch or relay contacts, make
certain that the polarity is correct (e.g. connect the ground pin of the Probot to
the ground pin of the controlling device). The output signal of the controlling
device must be TTL level compatible.
Содержание Probot 161403
Страница 1: ...D960R3 Probot TM Micro Fraction Collector User s Manual P N 161403 www lcpackings com ...
Страница 6: ...Table of Contents vi D960R3 User s Manual Probot This page intentionally left blank ...
Страница 10: ...Instructions for Returning Instruments x D960R3 User s Manual Probot This page intentionally left blank ...
Страница 14: ...CE Declaration xiv D960R3 User s Manual Probot This page intentionally left blank ...
Страница 68: ...Installation and Getting Started 2 48 D960R3 User s Manual Probot This page intentionally left blank ...
Страница 102: ...Appendix C C 6 D960R3 User s Manual Probot This page intentionally left blank ...
Страница 112: ...Appendix D D 10 D960R3 User s Manual Probot This page intentionally left blank ...