Installation and Getting Started
2-46
D960R3
User’s Manual Probot
The following section describes the procedure how to measure the dimensions of
your sample carrier by using the manual control of the Probot.
a)
Using one of the remote control options, move the table so that the needle is
positioned exactly above to the center of the first vessel.
b)
Choose the
Internal
button and read the current coordinates.
c)
Choose the
External
button, and then move the table along the X axis so that
the needle is positioned exactly above to the center of the last vessel.
d)
Read the coordinates again and divide the distance between the first and last
vessel by the number of vessels in that row/column minus 1. Enter this value
in the
SampleCarrierEditor
.
e)
Repeat step a) to d) with the Y axis.
f)
Repeat the steps above for any target placed on the table.