User’s Manual Probot
D960R3
A-1
APPENDIX A
Dosage Module Purge Routine
A.1
Overview
The dosage module (optional) is provided to add a reagent solvent (e.g. a matrix
solution) to the mobile phase flow. The figures presented below are provided to
describe the step-by-step procedure of how to remove air from the syringe and
how to purge the dosage module. For a detailed description of the software
control of the dosage module, refer to the
µ
Carrier software user’s manual.
Note: All fluidics pathways of the dosage module (e.g. the rotary valve, the
syringe and all tubing) must be free of any air bubbles and must be properly
purged.
A.1.1
The Step-By-Step Purge Routine
(FIGURE 2-5, Page 2-10) presents the dosage module with the syringe installed
in the syringe drive. FIGURE A-1 presents a schematic of this normal
configuration.
Wast
e
Fi
ll
Aux
(a
)
Reagent
Vials
Syringe Pump
Dosage Module
PUMP
FILL
PURGE
Pump
Fill
Rotary Valve
Needle
Tee
µ-Tee
OR
AUX
a) Wash Solvent
FIGURE A-1 Normal Configuration of the Dosage Module
Содержание Probot 161403
Страница 1: ...D960R3 Probot TM Micro Fraction Collector User s Manual P N 161403 www lcpackings com ...
Страница 6: ...Table of Contents vi D960R3 User s Manual Probot This page intentionally left blank ...
Страница 10: ...Instructions for Returning Instruments x D960R3 User s Manual Probot This page intentionally left blank ...
Страница 14: ...CE Declaration xiv D960R3 User s Manual Probot This page intentionally left blank ...
Страница 68: ...Installation and Getting Started 2 48 D960R3 User s Manual Probot This page intentionally left blank ...
Страница 102: ...Appendix C C 6 D960R3 User s Manual Probot This page intentionally left blank ...
Страница 112: ...Appendix D D 10 D960R3 User s Manual Probot This page intentionally left blank ...