48
Hinweise zur
elektromagnetischen
Verträglichkeit (EMV)
Electromagnetic
Compatibility
|
(EMC)
Information
ƃƚƐƗƐƝƘƯ
Ɵƞ
ƭƛƕƚƢƠƞƜƐƓƝƘƢƝƞƙ
ơƞƒƜƕơƢƘƜƞơƢƘ
(
ƍżƁ
)
Table 1
Guidance and manufacturer’s declaration – electromagnetic emissions
HAMOU
®
ENDOMAT
®
model
26
3310 20-1 is intended for use in the electromagnetic environment specified below.
The customer or user of the HAMOU
®
ENDOMAT
®
model
26
3310 20-1 should ensure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
HAMOU
®
ENDOMAT
®
model
26
3310 20-1 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B
The HAMOU
®
ENDOMAT
®
model
26
3310 20-1 is suitable for use in all establishments
including domestic establishments and those directly connected to the public low-
voltage power supply network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Complies
Accessories and cables which have been shown to comply with IEC 60601-1-2:
Type
Shielded
Length [m]
Ferrite
Used for
20
0103 30
Yes
3
No
Footswitch model
20
0103 30
20
0900 30
Yes
4
No
SCBcom model
20
0900 30
PA
No
> 3
No
Potential equalization
Power cord
No
3
No
Power supply