Calibration
WTX110
A4774-1.0
HBM: public
171
S
Three 1000kg load cells with sensitivity of 1.99995mV/V, 2.00005mV/V and
2.00000mV/V respectively are used.
S
The maximum load is 1500kg; the scale division is 0.5kg.
S
The weight of the empty tank is 600kg.
Zero point
Load point
Fig. 11.3
Example of calibration of a tank weigher
Menu
Key
Function/Secondary selection
Select group 1-9
Press 2 x for group
2 Calibration
→ Linearization points present
Fixpoints active!
Linearization entries have already been made;
they may influence the calibration.