Calibration
WTX110
A4774-1.0
HBM: public
161
Tare balancing and tare input values are carried over into the higher range
when load is applied, and are automatically rounded to the higher scale
interval.
The multi-range weighing machine permits configuration of a single SG base
unit for both low and high loads.
When configuring the multi-range weighing machine, all ranges and scale
intervals can be freely selected; only the number of decimal places has to be
the same for all ranges. If 0.5 kg and 1 kg scale intervals are used, for
example, 0.5 and 1.0 must be entered as the intervals. Additionally, the
maximum resolution of the load cells being used (e.g. 3000d) must not be
exceeded in any of the set ranges.
11.2
Multi-interval weighing machine
The single-range balance can be set as a multi-interval weighing machine with
two or three different weighing ranges, or scale intervals.
Example of a three-interval scale with:
Lowest weighing range: 0 kg to 1500 kg / 0.5 kg
Middle weighing range: 1500 kg to 3000 kg / 1.0 kg
Highest weighing range: 3000 kg to 6000 kg / 2.0 kg
The weighing range, or scale interval, is automatically switched dependent on
the applied load both when loading and when removing the load.
Содержание WTX110
Страница 1: ...Operating Manual Bedienungsanleitung English Deutsch WTX110...
Страница 3: ...Operating Manual Bedienungsanleitung English Deutsch WTX110...
Страница 10: ...8 A4774 1 0 HBM public WTX110 17 Geo values 212...
Страница 55: ...Installation WTX110 A4774 1 0 HBM public 53 Fig 4 12 Positioning of ALI option on the WTX110 motherboard...
Страница 217: ...Operating Manual Bedienungsanleitung English Deutsch WTX110...
Страница 224: ...8 A4774 1 0 HBM public WTX110 17 Geowerte 217...
Страница 259: ...Installation WTX110 A4774 1 0 HBM public 43 Abb 4 6 Positionierung DIO Modul SIO auf der WTX110 Hauptplatine...
Страница 271: ...Installation WTX110 A4774 1 0 HBM public 55 Abb 4 12 Positionierung Alibi Speicher auf WTX110 Hauptplatine...