Calibration
WTX110
A4774-1.0
HBM: public
185
Function/Secondary selection
Key
Menu
+X Y=0
Linearization of the +X side
—X Y=0
Linearization of the -X side
X=0 +Y
Linearization of the +Y side
X=0 —Y
Linearization of the -Y side
Lin.Load(X=0 Y=0)?
N
Determine the weight without the effect of the
optical tilt sensor. The scale must be placed on a
straight level to do this.
Y
Yes
N
No
Load(X=0 Y=0) = Y
X= 0.754 Y= -0.110?
Display of the current X, Y angle
Mount load and confirm; continue with
measurement on straight level
Linear.(X=0 Y=0) ...
Measurement is performed.
Lin.Load
100.0
Display/input of linearization weight
X= 0.754 Y= -0.110?
Display of the current X, Y angle Position the
scale at the inclination to be calibrated (here +X).
Linear.(+X Y=0) ...
Measurement is performed.
Weight:
100.01
Display the weight value in 10 times higher
resolution (as a check of the scale)
Содержание WTX110
Страница 1: ...Operating Manual Bedienungsanleitung English Deutsch WTX110...
Страница 3: ...Operating Manual Bedienungsanleitung English Deutsch WTX110...
Страница 10: ...8 A4774 1 0 HBM public WTX110 17 Geo values 212...
Страница 55: ...Installation WTX110 A4774 1 0 HBM public 53 Fig 4 12 Positioning of ALI option on the WTX110 motherboard...
Страница 217: ...Operating Manual Bedienungsanleitung English Deutsch WTX110...
Страница 224: ...8 A4774 1 0 HBM public WTX110 17 Geowerte 217...
Страница 259: ...Installation WTX110 A4774 1 0 HBM public 43 Abb 4 6 Positionierung DIO Modul SIO auf der WTX110 Hauptplatine...
Страница 271: ...Installation WTX110 A4774 1 0 HBM public 55 Abb 4 12 Positionierung Alibi Speicher auf WTX110 Hauptplatine...