Calibration
164
A4774-1.0
HBM: public
WTX110
After calibrating the scale, the approval parameters can be saved to the
EEPROM of the A/D converter, where they are retained even in the event of a
power outage. The calibration data can be saved on the ADM motherboard
with jumper W1. In the factory settings W1 is always set to calibration enabled!
Calibration data saved: Jumper set
Enable calibrating and saving: Jumper in
this position or completely removed
Information
Not removing the W1 jumper completely, or only removing it for replugging and
proceeding with care when moving the jumper is recommended to prevent the
jumper from being lost!