Calibration
WTX110
A4774-1.0
HBM: public
165
When setting the jumper, take care to ensure it is set on both pins to ensure
reliable calibration.
The position of the W1 jumper can be locked by a
weights and measures officer with thread and lead or a
seal: A description of calibration can be found in
section
!
In legal for trade applications, the covering and lid must be sealed by adhesive
labels so that the position of the calibration switch can no longer be changed.
→ If switch W1 saved in calibration data position
Calibration locked
Switch W1 not in calibration position, parameters
cannot be saved!
Access calibration without save option (e.g. for
checking)