Calibration
WTX110
A4774-1.0
HBM: public
159
11
Calibration
To access the
Calibration
group select the following keys:
Menu
Key
Function
Standard
Switch from
Standard
to version notice
WTX110V#.#
Display the current firmware version number
Call password entry within 3 seconds
Password
324
Enter the password
1 Service mode
Call
Service mode
Service:
Interface
Press twice to switch to the
Calibration
group
options
Service:
Calibration
11.1
Multiple-range
The A/D converter can be configured as a single, dual or triple-range scale.
The individual ranges differ in their maximum load and scale interval.
Example of three-range scale with a resolution of 3000 increments in each
weighing range:
Lowest weighing range: 0 to 1500kg / 0.5 kg
Middle weighing range: 0 to 3000kg / 1.0 kg
Highest weighing range: 0 to 6000kg / 2.0 kg
Содержание WTX110
Страница 1: ...Operating Manual Bedienungsanleitung English Deutsch WTX110...
Страница 3: ...Operating Manual Bedienungsanleitung English Deutsch WTX110...
Страница 10: ...8 A4774 1 0 HBM public WTX110 17 Geo values 212...
Страница 55: ...Installation WTX110 A4774 1 0 HBM public 53 Fig 4 12 Positioning of ALI option on the WTX110 motherboard...
Страница 217: ...Operating Manual Bedienungsanleitung English Deutsch WTX110...
Страница 224: ...8 A4774 1 0 HBM public WTX110 17 Geowerte 217...
Страница 259: ...Installation WTX110 A4774 1 0 HBM public 43 Abb 4 6 Positionierung DIO Modul SIO auf der WTX110 Hauptplatine...
Страница 271: ...Installation WTX110 A4774 1 0 HBM public 55 Abb 4 12 Positionierung Alibi Speicher auf WTX110 Hauptplatine...