background image

INSTALLATION INSTRUCTIONS

NI TB-2643

4 × 64 2-Wire or Dual 4 × 64 1-Wire Matrix Terminal Block

 

for the NI PXI-2532

Introduction 

The NI TB-2643 terminal block configures your NI PXI-2532 as a 4 

×

 64 

2-wire matrix or dual 4 

× 

64 1-wire matrix. The NI TB-2643 has ribbon 

cable headers to connect signals to the switch, and it provides optional 
isolation resistors to protect the reed relays from capacitive loads.

Refer to the 

NI Switches Getting Started Guide

 to determine when to install 

the terminal block.

Make sure you have the following:

NI TB-2643 terminal block

1/8 in. flathead and #1 Phillips screwdrivers

Eight 2 mm jumpers

Four 34 conductor, 28 AWG, .050 in. pitch ribbon cable assemblies 
(not included)

One 16 conductor, 28 AWG, .050 in. pitch ribbon cable assembly 
(not included)

Note 

Refer to the 

Accessories

 section for information about ordering the appropriate cable 

assemblies.

Connecting Ribbon Cables

To connect ribbon cables to the terminal block, refer to Figures 1 and 2 
while completing the following steps:

1.

Remove the top cover screw.

2.

Gently remove the top cover from the terminal block.

̭͈Ρ΅νιϋΠ͉ͅȂ඾ུࢊβȜΐ͜܄̞̳ͦ̀͘͘ȃ

Содержание TB-2643

Страница 1: ... determine when to install the terminal block Make sure you have the following NI TB 2643 terminal block 1 8 in flathead and 1 Phillips screwdrivers Eight 2 mm jumpers Four 34 conductor 28 AWG 050 in pitch ribbon cable assemblies not included One 16 conductor 28 AWG 050 in pitch ribbon cable assembly not included Note Refer to the Accessories section for information about ordering the appropriate ...

Страница 2: ...6 Connect each ribbon cable to the appropriate headers on the column connection boards and the module interface board 7 Reassemble the terminal block Note For information about protection resistance and matrix expansion refer to the NI Switches Help Figure 1 NI TB 2643 Terminal Block 1 Top Cover Screw 2 Top Cover 3 Strain relief Screws 4 Strain relief Bar 5 Column Connection Board Screws 6 Lower C...

Страница 3: ...connection cable for column expansion National Instruments 779325 01 16 Conductor 05 in pitch ribbon cable assemblies Samtec FFSD 08 01 N 34 Conductor 05 in pitch ribbon cable assemblies Samtec FFSD 17 01 N J4 3 2 1 0 Pin 1 3 2 1 0 LOWER COLUMN CONNECTION BOARD C16 C16 C0 C0 C15 C15 UPPER COLUMN CONNECTION BOARD C31 C31 C47 C47 C63 C48 C48 C32 C32 C63 J2 J3 J2 J3 3 3 2 2 1 1 0 0 Pin 1 J3 3 2 1 0 P...

Страница 4: ...stics All input characteristics are DC ACpk or a combination unless otherwise specified Maximum switching voltage Channel to channel 100 V Channel to ground 100 V CAT I Caution This module is rated for Measurement Category I and intended to carry signal voltages no greater than 100 V This module can withstand up to 500 V impulse voltage Do not use this module for connections to signals or for meas...

Отзывы: