Installation
52
A4774-1.0
HBM: public
WTX110
Parameters
Type
Low batt. (volts)
(flashing battery
symbol beginning at
about)
Power off (volts)
(shutdown at about)
Pb12
12 V lead
rechargeable
battery
11.5 V
11 V
Pb24
24 V lead
rechargeable
battery
23 V
22 V
Adjust
Any recharge
able battery
Adjustable
Adjustable
Line
Mains operation
-
-
If there is only a little remaining capacity (Low batt) the user is alerted by a
flashing battery icon in the top right corner of the display that the rechargeable
battery must be charged.
To prevent deep discharge of the connected rechargeable battery, the terminal
automatically switches off at the set voltage (Power off). If the terminal is
turned on again with the rechargeable battery deeply discharged, a message
appears indicating 'Low Battery: Power off' and the terminal switches off again.
4.17
Legal for trade data storage/alibi memory
If a WTX110 with the ALI option was ordered, the legal for trade memory is
completely pre-wired and plugged into the ASM slot. It is used for storing the
last 1,000,000 weighing processes.
K-WTX110-*-***
-
**-***-**-
ALI
-**
Содержание WTX110
Страница 1: ...Operating Manual Bedienungsanleitung English Deutsch WTX110...
Страница 3: ...Operating Manual Bedienungsanleitung English Deutsch WTX110...
Страница 10: ...8 A4774 1 0 HBM public WTX110 17 Geo values 212...
Страница 55: ...Installation WTX110 A4774 1 0 HBM public 53 Fig 4 12 Positioning of ALI option on the WTX110 motherboard...
Страница 217: ...Operating Manual Bedienungsanleitung English Deutsch WTX110...
Страница 224: ...8 A4774 1 0 HBM public WTX110 17 Geowerte 217...
Страница 259: ...Installation WTX110 A4774 1 0 HBM public 43 Abb 4 6 Positionierung DIO Modul SIO auf der WTX110 Hauptplatine...
Страница 271: ...Installation WTX110 A4774 1 0 HBM public 55 Abb 4 12 Positionierung Alibi Speicher auf WTX110 Hauptplatine...