Interference
210
A4774-1.0
HBM: public
WTX110
Remedy
Possible causes
Error message display
Out of Range
A/D converter over/undermodulating because:
Load cell incorrectly
connected
Check wiring
and measurement signal:
0–20 mV/AC (signal)
approx. 5 V/AC
(excitation)
Load cell faulty
Check load cell
Extreme overload on
scale
Unload scale
O v e r l o a d
Scale overloaded
Unload scale
--------
Zeroing or taring cannot
be performed because
the scale is in motion
Stabilize the scale
CPU not receiving data
from weighing interface
Check external and
internal wiring
U n d e r l o a d
Gross weight values less
than
*
20 d (below zero)
Load the scale
Set parameter 'Underload
20d' to N=Off
Power-up Motion
Power-up motion.
This message appears
immediately after power-
up if the scale does not
find a stable weight value
within the set power-up
zero range (±2 %,
±10 %).
Stabilize the scale
Содержание WTX110
Страница 1: ...Operating Manual Bedienungsanleitung English Deutsch WTX110...
Страница 3: ...Operating Manual Bedienungsanleitung English Deutsch WTX110...
Страница 10: ...8 A4774 1 0 HBM public WTX110 17 Geo values 212...
Страница 55: ...Installation WTX110 A4774 1 0 HBM public 53 Fig 4 12 Positioning of ALI option on the WTX110 motherboard...
Страница 217: ...Operating Manual Bedienungsanleitung English Deutsch WTX110...
Страница 224: ...8 A4774 1 0 HBM public WTX110 17 Geowerte 217...
Страница 259: ...Installation WTX110 A4774 1 0 HBM public 43 Abb 4 6 Positionierung DIO Modul SIO auf der WTX110 Hauptplatine...
Страница 271: ...Installation WTX110 A4774 1 0 HBM public 55 Abb 4 12 Positionierung Alibi Speicher auf WTX110 Hauptplatine...