Calibration
WTX110
A4774-1.0
HBM: public
175
The linearization software supports up to 6 correction points, though in practice
1 or 2 are almost always enough. The linearization points should be placed at
the points on the weight curve where the greatest variations occur.
Linearization points must be lower than the set maximum load. The lines
between adjacent linearization points are calculated as straight, meaning the
variations are interpolated in linear mode.
Menu
Key
Function/Secondary selection
Select group 1-9
Press 3 x for group
3 Linearization
→ Linearization points already present
Fixpoint 1:
999999
Display existing linearization points (1 to max. 6)
Delete existing linearization point
Calibrate Fixpt?
N
Lin.Signal
0.50000
Linearization signal is displayed
→ Clr key pressed in
Fixpoint n
step
Delete Fixpoint n? N
Y
Delete linearization point; continue to
New
Fixpoint?
step
N
Do not delete linearization point; continue
to next linearization point
Information
After deleting a linearization point, the remaining points are renumbered in
ascending order.
Содержание WTX110
Страница 1: ...Operating Manual Bedienungsanleitung English Deutsch WTX110...
Страница 3: ...Operating Manual Bedienungsanleitung English Deutsch WTX110...
Страница 10: ...8 A4774 1 0 HBM public WTX110 17 Geo values 212...
Страница 55: ...Installation WTX110 A4774 1 0 HBM public 53 Fig 4 12 Positioning of ALI option on the WTX110 motherboard...
Страница 217: ...Operating Manual Bedienungsanleitung English Deutsch WTX110...
Страница 224: ...8 A4774 1 0 HBM public WTX110 17 Geowerte 217...
Страница 259: ...Installation WTX110 A4774 1 0 HBM public 43 Abb 4 6 Positionierung DIO Modul SIO auf der WTX110 Hauptplatine...
Страница 271: ...Installation WTX110 A4774 1 0 HBM public 55 Abb 4 12 Positionierung Alibi Speicher auf WTX110 Hauptplatine...