Calibration
176
A4774-1.0
HBM: public
WTX110
→ After displaying the last linearization point, or if there are no points
New Fixpoint 1?
N
Y
Enter linearization point
N
Do not enter linearization point; back to
Select Group
Enter Fixpt.1
999999
Enter the weight to be linearized
Calibrate Fixpt?
N
Y
Calibrate linearization point
N
Enter linearization value
→ With
Calibrate Fixpt = Y
Linearization..
Measurement of the linearization signal
Menu
Key
Function/Secondary selection
Weight:
9999999
Display the linearized weight value with 10 times
higher resolution (as a check)
Lin.Signal1
0.50000
Enter a calculated linearization signal or import
the values from another A/D converter
Continue in
New Fixpoint 2?
step
A new linearization point can be inserted between existing points. Then all the
points are renumbered in ascending order.
11.7
Zero Adjust
For more information on the keys necessary to access this menu group, go to
the
.
In this menu group the zero point of the scale can be redefined. This becomes
necessary in particular when the scale has been calibrated using an auxiliary
frame. After removing the auxiliary frame, the final zero point is then defined.
The change in zero point shifts the weighing range accordingly.
Содержание WTX110
Страница 1: ...Operating Manual Bedienungsanleitung English Deutsch WTX110...
Страница 3: ...Operating Manual Bedienungsanleitung English Deutsch WTX110...
Страница 10: ...8 A4774 1 0 HBM public WTX110 17 Geo values 212...
Страница 55: ...Installation WTX110 A4774 1 0 HBM public 53 Fig 4 12 Positioning of ALI option on the WTX110 motherboard...
Страница 217: ...Operating Manual Bedienungsanleitung English Deutsch WTX110...
Страница 224: ...8 A4774 1 0 HBM public WTX110 17 Geowerte 217...
Страница 259: ...Installation WTX110 A4774 1 0 HBM public 43 Abb 4 6 Positionierung DIO Modul SIO auf der WTX110 Hauptplatine...
Страница 271: ...Installation WTX110 A4774 1 0 HBM public 55 Abb 4 12 Positionierung Alibi Speicher auf WTX110 Hauptplatine...