Calibration
180
A4774-1.0
HBM: public
WTX110
Function/Secondary selection
Key
Menu
Overload
9999
Overload display suppression (in D) can be set
here. If the scale exceeds the value (maximum
load + overload), the display is hidden.
Example:
Overload 9
On multiple-range scales, 1d corresponds to the
digit of the lowest weighing range.
Incline Switch
Off
OFF
No incline switch connected
In 0…In7
Discrete input
Sensor
Connection to an incline switch for
correction or switch off
For more information
NTEP approval?
N
Switch to the metrological properties according to
American requirements (National Type
Evaluation Program) as per Handbook 44.
N
Off (default)
Y
On
Must be set to
Y
(es) for applications in the USA,
Canada and Australia!
Underload 20d?
Y
Hide weight indicator if gross weight values less
than -20d (below zero)
N
Off
Y
On
For applications in accordance with OIML R76,
Underload 20d
must be set to
Y
With Taring?
Y
Select whether tare functions are disabled or
enabled.
N
Tare functions disabled
Y
Tare functions enabled (default)
For scales that only operate in Gross mode, the
tare functions can be disabled here so as to skip
cost-intensive checking of the tared scale during
legal verification.
Содержание WTX110
Страница 1: ...Operating Manual Bedienungsanleitung English Deutsch WTX110...
Страница 3: ...Operating Manual Bedienungsanleitung English Deutsch WTX110...
Страница 10: ...8 A4774 1 0 HBM public WTX110 17 Geo values 212...
Страница 55: ...Installation WTX110 A4774 1 0 HBM public 53 Fig 4 12 Positioning of ALI option on the WTX110 motherboard...
Страница 217: ...Operating Manual Bedienungsanleitung English Deutsch WTX110...
Страница 224: ...8 A4774 1 0 HBM public WTX110 17 Geowerte 217...
Страница 259: ...Installation WTX110 A4774 1 0 HBM public 43 Abb 4 6 Positionierung DIO Modul SIO auf der WTX110 Hauptplatine...
Страница 271: ...Installation WTX110 A4774 1 0 HBM public 55 Abb 4 12 Positionierung Alibi Speicher auf WTX110 Hauptplatine...