10-34
Section 10-9 - When There's Too Much Leakage Current...
Type CF Patient Leakage Current Test - ECG Leads
ECG lead 1
ECG lead 2
ECG lead 3
Tester
Neutral
Tester Polarity
Switch
Ground Switch
Max Value
Allowed
Value
Measured
OK?
Comments
Closed
Normal
Closed
0.01mA
Closed
Normal
Open
0.05mA
Closed
Reversed
Closed
0.01mA
Closed
Reversed
Open
0.05mA
Open
Normal
Closed
0.05mA
Open
Reversed
Closed
0.05mA
Tester
Neutral
Tester Polarity
Switch
Ground Switch
Max Value
Allowed
Value
Measured
OK?
Comments
Closed
Normal
Closed
0.01mA
Closed
Normal
Open
0.05mA
Closed
Reversed
Closed
0.01mA
Closed
Reversed
Open
0.05mA
Open
Normal
Closed
0.05mA
Open
Reversed
Closed
0.05mA
Tester
Neutral
Tester Polarity
Switch
Ground Switch
Max Value
Allowed
Value
Measured
OK?
Comments
Closed
Normal
Closed
0.01mA
Closed
Normal
Open
0.05mA
Closed
Reversed
Closed
0.01mA
Closed
Reversed
Open
0.05mA
Open
Normal
Closed
0.05mA
Open
Reversed
Closed
0.05mA