GE
DRAFT
V
OLUSON
™ P8 / V
OLUSON
™ P6
DIRECTION 5723243, R
EVISION
6
DRAFT (M
AY
23, 2018)
B
ASIC
S
ERVICE
M
ANUAL
Chapter 3 - Setup Instructions
3-49
3-7-7-1
UP-D897 / UP-D898MD - Printer Settings
1.) Call up the
‘Printer Preferences’
; operation see:
Section 3-7-7 "Adjustment of Printer Settings"
.
2.) Select the
LAYOUT
page and select:
-
Paper:
960x1280
-
Orientation:
Portrait
-
Interpolation Method:
Bilinear
3.) Select the
DENSITY ADJUST
page and select:
-
Gamma:
TONE2
-
Advanced =
0
; Dark =
0
; Light =
0
; Sharpness =
2
4.) Save the adjusted printer settings click APPLY and then OK.
Close window and exit System Setup with SAVE&EXIT.
5.) Assign the Printer to the remote keys
P1
,
P2
, and/or
P3
with the save button;
Section 3-7-8 "Remote Control Selection" on page 3-52
Figure 3-51 Density Adjust page