Chapter 5 - Components and Functions (Theory)
5-57
5-15-1-3
Rollback
This function offers the possibility to simply restore the previously used system configuration (rollback),
which was stored on R:\ during “FMI from DVD”.
1.) Turn system OFF and then back ON.
2.) As soon as the “Boot Screen” appears (see:
), press the
PG DN
[
Arrow down] key on the keyboard until the
Rollback
item is highlighted, then press
ENTER
.
3.) When the following WARNING message appears, press the
Arrow left] button to highlight OK
and then press
ENTER
.
After performing the rollback, the system reboots.
The Ultrasound Application is started and finally the 2D screen is displayed on the monitor.
CAUTION
After 3 sec. without pressing any key, the system will boot-up in windows (= Voluson item).
If you missed step 2, retry again with
CTRL
+
ALT
+
DEL
.
Figure 5-32 Warning message
select [OK]