Chapter 10 - Care & Maintenance
10-35
Type BF Patient Leakage Current Test - Probes (This test should be done for all probe)
Type CF Patient Auxiliary Current Test - ECG Lead to Probes
Type BF Patient Leakage Current Test - Mains to Probe
Probe (This test should be done for all probes)
Type CF Patient Leakage Current Test - Mains to ECG leads
ECG leads connected all together
Tester
Neutral
Tester Polarity
Switch
Ground Switch
Max Value
Allowed
Value
Measured
OK?
Comments
Closed
Normal
Closed
0.1mA
Closed
Normal
Open
0.5mA
Closed
Reversed
Closed
0.1mA
Closed
Reversed
Open
0.5mA
Open
Normal
Closed
0.5mA
Open
Reversed
Closed
0.5mA
Tester
Neutral
Tester Polarity
Switch
Ground Switch
Max Value
Allowed
Value
Measured
OK?
Comments
Closed
Normal
Closed
0.01mA
Closed
Normal
Open
0.05mA
Closed
Reversed
Closed
0.01mA
Closed
Reversed
Open
0.05mA
Open
Normal
Closed
0.05mA
Open
Reversed
Closed
0.05mA
Tester
Neutral
Tester Polarity
Switch
Ground Switch
Max Value
Allowed
Value
Measured
OK?
Comments
Closed
Normal
Closed
5mA
Closed
Reversed
Closed
5mA
Tester
Neutral
Tester Polarity
Switch
Ground Switch
Max Value
Allowed
Value
Measured
OK?
Comments
Closed
Normal
Closed
0.05mA
Closed
Reversed
Closed
0.05mA