Chapter 10 - Care & Maintenance
10-23
10-8-5-3
Data Sheet for Enclosure Leakage Current
The test passes when all readings measure less than the value shown in
Record all data on the Ultrasound Inspection Certificate.
10-8-6
Earth Leakage Current Test
10-8-6-1
Definition
This test measures the current that would flow to a grounded person who touched accessible
protectively-earthed parts of equipment if the ground wire should break.
The meter is connected from accessible protectively-earthed parts of the equipment to ground.
Measurements should be made with power line polarity normal and reversed, and with the neutral open
and closed Record the highest reading.
Table 10-23 Typical Data Sheet for Enclosure Leakage Current
Tester
Neutral
Tester Polarity
Switch
Protective Earth
Switch
Test 1
Wheel
Test 2
LCD
Optional
Test 3
Optional
Test 4
Enter Name of tested peripheral here:
Closed
Normal
Closed
Closed
Normal
Open
Closed
Reversed
Closed
Closed
Reversed
Open
Open
Normal
Closed
Open
Reversed
Closed
CAUTION
!! CAUTION:
Equipment damage possibility, Never switch the Polarity and the status of Neutral when the unit
is powered ON. Be sure to turn the unit power OFF before switching them using the POLARITY
switch and/or the NEUTRAL switch. Otherwise, the unit may be damaged.