5-62
Section 5-17 - LINUX - Voluson Maintenance System
5-17-3
Before Recovering or Initializing the Hard disk
1.) Boot up the system, connect a probe and start any user program.
2.) Press the
UTILITIES
key on the control console.
3.) In the Utilities menu, select
SETUP
button to invoke the setup desktop on the screen.
4.) On the right side of the screen select
ADMINISTRATION
:
- Click the OPTIONS tab to see which options are currently installed in the system.
Please print out this page or write down the the state of the option (P = Permanent, I =
inactive) and the “Permanent Key”. If this was not possible, contacting the OLC will be
necessary.
- Click the
Service
tab, enter the password SHE and then click
Accept
. In the displayed
Service Tools window click the
Keyboard Layout
button to check the default input language
setting.
- Click the
System Info
tab and check Software Version and date of the “Factory Settings”
(see:
Section Figure 7-1 "System Setup - Administration - SYSTEM INFO page (e.g., VP8)"
5.) On the right side of the screen select
BACKUP
:
- Click the
SYSTEM CONFIGURATION
tab and then click on the SAVE button of the “Full
System Configuration” group. Select destination, enter description and then click on NEXT.
For more details about “Full Backup” see: Section 4-5-3 on page 4-37.
- Click the
IMAGE ARCHIVE
tab and then click on SAVE.
Select destination, enter the description and then click on NEXT.
For more details about saving the Patient Image Archive, see:
Section 4-4-1 "Patient Archive
(Image Management)" on page 4-9
CAUTION
If the system is still responsive (system starts up without problems and the application software
is running as usual, try to print out or write down System Setup and Measure Setup settings;
because they could be lost. Even if most of these settings can be restored after the installation,
it is strongly recommend to record them before starting the procedure.
Disconnecting ALL external USB devices (except DVD/CD+R/RW drive) is NECESSARY.
Re-adjusting those settings and re-installation of any previously attached printer has to be done
after the recovery procedure
CAUTION
If the system is suspected to be infected with a virus, it is highly recommended to disconnect
the network cable before starting the system.
NOTICE
If the currently installed software has to be upgraded to a newer version, a new software specific
“Permanent key is required. (Calculated in OKOS - Option key Order System;
http://3.187.187.9/
OKOS
. For detailed description see . If you do not have permission to access OKOS, contact the On
Line Center to get the mecessary key.