Section 5-15 - Boot Screen Functions
Section 5-15
Boot Screen Functions
5-15-1
Overview
NOTE:
Prior to the “Boot Screen” ( see figure below), the BIOS version is shown on the monitor:
Following LINUX supported functions are available as soon as the “Boot screen” appears:
5-15-1-1
Voluson
The System will boot-up in windows. The Ultrasound Application is started.
For details refer to
Section 3-6-2-3 "During a normal boot, you may observe" on page 3-28
5-15-1-2
GE-Service
The
Voluson Maintenance System
functions are secured by service password (class C).
1.) Turn system OFF and then back ON.
2.) As soon as the “Boot Screen” appears (see:
above), press the
PG DN
[ Arrow
down] key on the keyboard until the
GE Service
item is highlighted, then press
ENTER
.
3.) Enter the Password
Rudi
(case sensitive) and confirm with
ENTER
. The system boots into
LINUX.
Figure 5-31 Boot screen
NOTICE
!! NOTICE:
After 3 sec. without pressing any key, the system will boot-up in windows (=
Voluson
item). If you
missed selection, retry again with Ctrl+Alt+Del.
CAUTION
After 3 sec. without pressing any key, the system will boot-up in windows (= Voluson item).
If you missed step 2, retry again with
CTRL
+
ALT
+
DEL
.
Voluson
GE-Service
Rolback
Memtest(Memory Check)