Chapter 3 - Setup Instructions
3-29
3-6-3
Power Off / Shutdown
3-6-3-1
Ultrasound unit Shutdown
1.) If not already in read mode, freeze the image.
2.) Press the
ON/OFF
button (see:
) on the control panel. Following dialog appears.
3.) Select the SHUTDOWN button. The system performs an automatic full shutdown sequence.
NOTE:
Full shut down is also performed when pressing the ON/OFF button on the control panel twice.
4.) Switch OFF the Circuit at the rear of the system.
NOTICE
!! NOTICE:
After turning off a system, wait at least 10 seconds before turning it on again.
The system may not be able to boot if power is recycled too quickly.
Figure 3-17 Shutdown, Restart or Cancel