GE
RAFT
V
OLUSON
™ P8 / V
OLUSON
™ P6
DIRECTION 5723243, R
EVISION
6
DRAFT (M
AY
23, 2018)
B
ASIC
S
ERVICE
M
ANUAL
7-52
Section 7-10 - Troubleshooting Trees, Instructions and Tech Tips
7-10-8
Tech Tips
Use
below to access Voluson™ P8 / Voluson™ P6 (troubleshooting) Technical Tips listed.
Each entry in the table will hyper-link to the issue, cause and solution.
7-10-8-1
Storing SonoView images to Voluson™ P8 / Voluson™ P6
Procedure:
1.) Connect external hard disk or insert CD/DVD on the Voluson™ P8 / Voluson™ P6.
2.) Open Archive --> select "Data Transfer" --> "Import". The "Open File" window appears
3.) Select Drive in the “Look In” pull down menu.
4.) In the “File of type” field change file type to “V730 Format (v730.mdb)”.
5.) Browse for the folder where SonoView Backup was stored.
6.) Select file "v730.mdb" and then click OPEN.
Table 7-9
Technical Tips
Sub-section
Description
Page Number
Storing SonoView images to Voluson™ P8 / Voluson™ P6
Issue:
Storing SonoView images from Voluson
TM
730/Expert/Pro/ProV to Voluson™ P8 / Voluson™ P6
Cause:
Archive is different (no SonoView on Voluson™ P8 / Voluson™ P6)
Solution:
1.) Perform SonoView Backup on Voluson
TM
730/Expert/Pro/ProV to external hard disk (USB) or DVD
2.) Import File "V730.mdb" from external hard disk (USB) or CD/DVD on Voluson™ P8 / Voluson™ P6
Figure 7-44 Open File