Chapter 5 - Components and Functions (Theory)
5-79
5-17-7-6
CollectLogs
Voluson Log files can be stored on an USB-stick. The stored zip file includes the current system state
(as “Full Backup”), Process Logger files, Auto tester files, SMART logs, dump files, etc.
1.) Connect an USB-stick.
2.) In the “Service Tools Menu” select [CollectLogs] and then press Enter.
3.) In the displaye screen select your USB stick and then press the Space key. An asterisk* appears
between teh brackets.
4.) Click
OK
to save the log files.
5-17-7-7
CheckLnkSpeed
This checks link speed communication between Beamformer and BackEnd Processor. x4 ==> Passed
Figure 5-54 Select log destination by pressing Space