GE
V
OLUSON
i / V
OLUSON
e
D
IRECTION
KTI106052, R
EVISION
10
S
ERVICE
M
ANUAL
Chapter 4 - Functional Checks
4-39
Section 4-7
Peripheral Checks
Check that peripherals work as described below:
4-7-1
Power Supply Adjustment
There are no adjustments on the power supplies. The DC Power is self-regulated. If a voltage is outside
the specified range, it means that something is wrong, either with the power supply itself or with a
component connected to that specific power outlet.
Table 4-12 Peripheral Checks
Step
Task to do
Expected Result(s)
1
Press the
FREEZE
key.
Stop image acquisition.
2
Press the
P1
,
P2
and/or
P3
key on the Control Panel.
The image displayed on the screen is printed, saved or
exported, depending on the key assignment configuration