GE
V
OLUSON
i / V
OLUSON
e
D
IRECTION
KTI106052, R
EVISION
10
S
ERVICE
M
ANUAL
10-4
Section 10-3 - Maintenance Task Schedule
Probe/Phantom Checks
Check axial and lateral resolution (see Basic
User Manual Technical specifications).
Check Gain and TGC changes, vary the
focus and check reaction on screen.
NOTE!
The use of a Phantom is not required during
Preventive Maintenance. Customer may
use it as part of their Quality Assurance
Program tests
.
•
Also after corrective
maintenance or as required by
your facilities QA program.
Functional Checks
of all probes
Section 10-
5-2 on page 10-6
NOTE! The use of a Phantom is not required
during Preventive Maintenance. Customer
may use it as part of their Quality Assurance
Program tests.
•
Also after corrective
maintenance or as required by
your facilities QA program.
Table 10-2 Customer Care Schedule
Item
Service at Indicated Time
D
W
M
A
Notes
Содержание Voluson i BT06
Страница 2: ......
Страница 11: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL ix ZH CN KO...
Страница 44: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL xlii Table of Contents...
Страница 514: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL IV Index...
Страница 515: ......