Section 3-5 - Connection of Auxiliary Devices
3-5-7
Connecting the DVD+R/RW Drive
3-5-7-1
Connection directly to the Voluson i / Voluson e system
The DVD+R/RW Drive can be connected to the USB ports on the rear connector panel.
NOTE:
The connection Scheme for DVD Drive Samsung, De-Lock and Teac is similary the same.
Figure 3-60 Connection Scheme - DVD+R/RW Drive
CAUTION
!! CAUTION:
When connecting directly to the system, it is necessary to use additional AC/DC power supply (supplied
with the drive).
As a safety requirement, the AC/DC must be medical grade (conforming 60601-1
standard) or the safety must be ensured by the use of a medical grade isolation transformer.
NOTICE
!! NOTICE:
Auxiliary equipment with direct mains connection requires galvanic separation e.g. by use of a medical
grade isolation transformer.
Power Supply
power
socket
USB cable
isolation transformer
NOTICE
:
Auxiliary equipment with direct mains
connection requires galvanic
separation e.g. by use of a medical
grade isolation transformer.