Chapter 5 - Components and Functions (Theory)
5-21
Section 5-4
Back End Processor
5-4-1
Introduction
The Back End Processor (BEP) supports the operation of the
Voluson i / Voluson e ultrasound unit and
is the main controller for the system.
The various components that comprise the Voluson i / Voluson e BEP are described in the following
sections:
•
ETX SBC Central Processing Unit (CPU) on page 5-22
.
•
ETX Base Board on page 5-22
.
•
Bluetooth Module and Antenna on page 5-23
.
•
Hard Disk on page 5-24
.
•
15" LCD Monitor on page 5-27
Figure 5-5 Voluson i / Voluson e Back End Processor - Block Diagram
Front End
GPM (RPM)
Back End
Baseboard
GPP (RPS)
SMBus
Rx-MUX
192 to 64
Probe
Connector
192 El.
RBF (1)
32 x (Tx+Rx)
Control
Control
RBF (2)
32 x (Tx+Rx)
ETX
SBC
Pentium M
1.4GHz
HV-Supply
LV-Supply Frontend
LV-Supply Backend
Battery
RPC
Controller & Monitoring
Motor
Power
Amplifier
8x
BFIC
2
nd
Monitor
HDD (80GB)
Supply
Supply
AC Adapter
Interface for
Docking Station ..
PMC - Power
Management
Controller
SMB
us
+ Charger
System Supply
Tx-MUX
192 to 64
Buffer
DC
Source-
Selector
SMB
us
Bottom Assembly
Motor
Controller
5%,
PCI
PCMCIA
LAN 100 MBit
VGA
2x USB2.0
Содержание Voluson i BT06
Страница 2: ......
Страница 11: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL ix ZH CN KO...
Страница 44: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL xlii Table of Contents...
Страница 514: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL IV Index...
Страница 515: ......