Chapter 3 - Setup Instructions
3-73
3-6-4
Internal Lithium Ion Battery
One battery is standard delivered with the Voluson i / Voluson e. This lithium ion battery provides power
when an AC power source is not available. You can expect one hour of battery life with a fully charged
battery.
When the system is running, a status icon is displayed in the system Status bar to indicate the current
battery status.
NOTE:
To check capacity of the battery, move the cursor over the
Battery Icon
in the status area.
When shutting down the system, leave the main power cable connected to remain in standby and
therefore battery fully charged.
Table 3-11 Battery Status
Icon
Status Description
AC Power is on; no battery present
AC Power is on; battery is full
Battery in use; battery almost empty
Battery in use; fully charged
Battery in use; partly charged
AC Power is on; battery is charging
Figure 3-91 Battery Capacity check
NOTICE
!! NOTICE:
If the system is left in the OFF condition (system unplugged from the wall outlet) for an extended period
of time (3 to 5 days or more), the system may not boot up or may beep when turned ON. If this occurs,
the battery needs to be recharged by means of the internal or the external (optional) battery charger.
Operation see:
Section 3-6-4-1 "Charging the Battery" on page 3-74
.
CAUTION
!! CAUTION:
If the Voluson i / Voluson e portable ultrasound system is not being used on a monthly basis,
the battery needs to be removed during lengthy non-use period.
Содержание Voluson i BT06
Страница 2: ......
Страница 11: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL ix ZH CN KO...
Страница 44: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL xlii Table of Contents...
Страница 514: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL IV Index...
Страница 515: ......