CRI: mismatch in steering angle header info
restart the system
CtrlMvGrpRepresentations::isInWriteMode not supported for
mode nr xx
restart the system
CW Archive::not implemented
restart the system
CW quant TX Frequ out of range
restart the system
CW_BaseLinePos Out Of Range
press ok and save this user-setting once again
CW_Gain Out Of Range
press ok and save this user-setting once again
CW-hardware doesn't support
pencil probe + CW-Hardware not available->HW problem
CW-HW-PRF == 0
restart the system
CW-TxFrequency == 0
restart the system
D:\\SERVICE\\ConfIndexFile is corrupt
restart the system
Data are not saved to Sonoview or sent to DICOM Server!
restart the system
Datatype misalignment
restart the system
dDistanceFocusReferenceBF_ToProbeSurface out of valid
range
restart the system
Delete error (Backup data could not be deleted.)
check storage destination of Full Backup
(e.g., DVD/CD, insufficient rights on target Network drive,
write protection on USB-drive)
Density Out Of Range
press ok and save this user-setting once again
Different software error (Backup data was made on another
system with a different software version.)
This backup cannot be restored by the user.
Disc is full!
use a new CDR for writing data to CD
Display:Rect Region fails
restart the system
Division by zero
restart the system
done is low!
restart the system
Doppler-Txfrequency out of range
restart the system
Downgrade error (Backup data was made with a software
version higher than the installed version.)
load appropriate backup for installed version
DSP - Gamma Corr - Load Data Timeout
restart the system
DSP - HilbertCoeff - Load Data Timeout
restart the system
DSP - Low Pass Coeff - Load Data Timeout
restart the system
DSP - SetFFT_Para - Load Data Timeout
restart the system
DSP - SetWMF_Koeff - Load Data Timeout
restart the system
DSP/MSE:Hanning-Window Load Data Timeout
restart the system
ECG_Consumer : memory allocation failed
restart the system
Error Messages
Actions
Содержание Voluson i BT06
Страница 2: ......
Страница 11: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL ix ZH CN KO...
Страница 44: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL xlii Table of Contents...
Страница 514: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL IV Index...
Страница 515: ......