Section 5-5 - Power Supply Unit
Section 5-5
Power Supply Unit
5-5-1
Introduction
The Power Supply Unit (GPP) provides the Voluson i / Voluson e system with both Low Voltage and
High Voltage power, and also enables recharging of the battery.
The GPP Unit controls following main functions:
-
Input switching stage – selects between the external DC source or internal battery
-
Back End Low Voltage power supply (LVPS)
-
Standby power supply
-
Front End Low Voltage power supply (LVPS)
-
Front End High Voltage power supply (HVPS)
-
Battery charger
-
PS ON/OFF control
-
Micro controller and circuitry (used for monitoring, charger control and auxiliary functions).
The PS communicates with the host CPU via the SMBus.
-
4D motor power - amplifier
5-5-2
Power Supply Unit Components
As shown in
Figure 5-10
below, the Power Supply Unit comprises the following components:
•
Rechargeable Battery
•
PS Controller and Monitor
•
DC Source Selector
•
Battery Charger
•
High Voltage Power Supply (HVPS)
•
Low Voltage Power Supply (LVPS)
Figure 5-10 Voluson i / Voluson e Power Supply Unit Components - Block Diagram